Differential Scanning Calorimetry

What is differential scanning calorimetry analysis?

Differential Scanning Calorimetry (DSC) is a primary technique in thermal analysis. It is instrumental in detecting endothermic and exothermic transitions, such as determining transformation temperatures and enthalpy changes in solids and liquids as a function of temperature.

DSC differs from traditional calorimetry, where a sample is placed in an isolated chamber to monitor its heat absorption and release in an isothermal experiment. Instead, DSC is a dynamic process.

In a typical calorimeter, a sample is situated within an isolated chamber, and it is heated with a known quantity of heat. The temperature difference between the sample and the surrounding medium provides information about the heat capacity of the sample, along with insights into the heat release and absorption by the sample. In contrast, the differential scanning technique employs both a sample and a reference material subjected to identical conditions, and their signals are directly subtracted from each other for analysis.

Where is Differential Scanning Calorimetry used?

DSC analysis finds application in a wide array of industries and is also used for fundamental research in academic settings. Its uses span diverse areas, including the determination of glass transition temperatures, examination of chemical reactions, and the study of melting and crystallization behaviors.

Furthermore, DSC is employed to investigate the effects of additives, fillers, and material processing. The unique shape of individual DSC curves is also harnessed for quality control purposes.

What is the difference between DTA and DSC?

The key distinction between DSC and DTA lies in their methodologies. DSC measures the difference in heat flow, whereas DTA assesses temperature variances between a reference sample and a sample of interest.

What is the difference between a heat flux DSC-analysis and a power-compensated DSC-analysis?

DSC devices are constructed based on two fundamental measuring principles: the heat-flux principle and the power-compensated DSC principle.

In the Heat-Flux DSC approach, changes in heat flow are determined by integrating the ΔTref (temperature difference reference) curve. In this type of DSC analysis, a sample and a reference crucible are positioned on a sample holder equipped with integrated temperature sensors to monitor the crucibles’ temperatures. This assembly is located within a temperature-controlled furnace. What distinguishes this DSC design from the conventional one is the vertical arrangement of planar temperature sensors surrounding a planar heater. This configuration results in a highly compact, lightweight, and low heat capacity structure while maintaining the full functionality of a DSC furnace.

In power-compensated DSC analysis, the sample and reference crucibles are physically separated. The temperature of both chambers is then controlled in a way that ensures both sides maintain the same temperature at all times. Instead of monitoring the temperature difference between the two crucibles, this method records the electrical power required to achieve and sustain this equilibrium.

What are the standards for differential scanning calorimetry?

Certainly, here are the rewritten descriptions for the mentioned standards:

  1. DIN 53765: This standard pertains to the testing of plastics and elastomers through thermal analysis, specifically using the DSC (Differential Scanning Calorimetry) method.
  2. DIN 51007:2019-04: This standard, updated in 2019, covers the general principles of thermal analysis, encompassing both Differential Thermal Analysis (DTA) and Differential Scanning Calorimetry (DSC).
  3. DIN EN 6041:2018-03: Part of the aerospace series, this standard involves the analysis of non-metallic materials (uncured) using Differential Scanning Calorimetry (DSC). It is available in both German and English versions.
  4. DIN EN 728: This standard relates to plastics piping and ducting systems, specifically the determination of oxidation induction time for polyolefin pipes and fittings. The German version corresponds to EN 728:1997.
  5. DIN EN ISO 11357-1, -2, -3, -4, and -6: These standards focus on plastics and employ Differential Scanning Calorimetry (DSC). Part 1 addresses general principles, and the documents are available in both German and English versions.
  6. ASTM E 126: This standard outlines the test method for inspecting, calibrating, and verifying ASTM hydrometers.
  7. ASTM E 1356: ASTM E 1356 establishes the test method for determining glass transition temperatures through Differential Scanning Calorimetry (DSC).